1. The instrument adopts waveform detection and Fourier transform to monitor the discharge current at any time, when the discharge current is monitored, immediately record the high voltage at the moment and cut off the high voltage power supply within 0.2s. The test result has high precision and is easy to realize automatic measurement.
2. Full touch LCD screen, large full graphical interface, every process is very clear. Operators do not need additional professional training to use it. The whole process can be tested with a touch.
3. The instrument is easy to operate, and the measurement of data is completed automatically under the control of microprocessor. No human intervention required.
4. Integrated model, with high voltage power supply, easy to field test, reduce field wiring.
5. High voltage protection: short circuit, breakdown or high voltage current fluctuation of the sample, can cut off the output in short circuit mode at high speed.
6. Low voltage protection: misconnect 380V, power fluctuation or sudden power outage, start protection, will not cause overvoltage.
7. Anti-misoperation: real-time monitoring of voltage and current; High/low voltage of terminal; Slow pressure boost, rapid depressurization, sound alarm.
8. Seismic performance: the instrument adopts unique seismic design, which can withstand strong long-distance transportation vibration and turbulence without damage.
9. High voltage cable: high voltage resistant insulated wire, but for accurate measurement, please use in the air.
Technical Indicators
1. Output voltage 0-30kV (effective value); Output current 0-25mA
2. Detection discharge mode: waveform detection and Fourier transform mode
3 accuracy: voltage: ± (reading ×2%+0.1) Current: ± (reading ×2%+0.1)
4. High pressure cut-off time: less than 0.2 seconds
Voltage resolution: 0.001kV current resolution: 0.001mA
6. Input power: 180V ~ 270VAC, 50Hz/60Hz±1%, mains or generator power supply
7. Computer interface: standard RS232 interface
8. Printer: Thermal micro printer
9. Ambient temperature: -10ºC ~ 50ºC Relative humidity: <90%
Product picture